Connection of Probes
The testing transmitters (test coils, scanning coils and hand-held rotating heads etc.) must be connected via a coil cable, possibly interconnecting an adapter at Socket 3 (Item 0.14) "test coil".
The testing transmitters belonging to the DEFECTOTEST system are provided with a matching connecting plug. Please note the following detailed information if it is intended to connect other special-purpose testing transmitters, possibly self-manufactured testing transmitters.
Socket 3 is a socket in accordance with Mil C 26 482, type PT 02 A 14-14S.
The matching mating plug has the designation PT 06E 14-15 P (SR).
Socket 3 has the following pin assignment:


Tester (Socket PT O02 A-14-155)

Test Coil (Plug PT O6A-14-15P (SR))